Acts Online
GT Shield

Legal Metrology Act, 2014 (Act No. 9 of 2014)

Regulations

Legal Metrology Regulations, 2017

Part XI : Measuring instruments and containers used for prescribed purposes

102. Pipettes

 

(1) Except as provided in this regulation, a pipette must conform to the applicable requirements of regulations 98 and 100.

 

(2)
(a) A pipette may have any value of maximum denomination not exceeding 5 L, provided that such denomination meets the requirements of paragraph (d) of this subregulation.
(b) A pipette may be provided with an enlarged section upon which no graduation lines must appear.
(c) A pipette may be provided with a narrowed top section upon which no graduation lines must appear.
(d) The value of the graduations of a pipette must be 1 x 10n, 2 x 10n or 5 x 10n of 1 mL, where the exponent "n" is a positive or negative whole number or zero, provided that, where the total denomination exceeds 1 L, the graduations may be figured in submultiples of 1 L.

 

(3) The errors allowed on a pipette are one-half of the allowances in Table XI, in the Annexure.